Laboratory centrifuge

Last updated
Laboratory centrifuge
Tabletop centrifuge.jpg
A Hettich tabletop laboratory centrifuge
Uses Separation
Related items Gas centrifuge
Ultracentrifuge

A laboratory centrifuge is a piece of laboratory equipment, driven by a motor, which spins liquid samples at high speed. There are various types of centrifuges, depending on the size and the sample capacity. [1]

Contents

Like all other centrifuges, laboratory centrifuges work by the sedimentation principle, where the centripetal acceleration is used to separate substances of greater and lesser density.

Types

Laboratory centrifuge Centrifuga Hermle 2.jpg
Laboratory centrifuge

There are various types of centrifugation:

There are different types of laboratory centrifuges:

Microcentrifuges
devices for small tubes from 0.2 ml to 2.0 ml (micro tubes), up to 96 well-plates, compact design, small footprint; up to 30,000 g
Clinical centrifuges
moderate-speed devices used for clinical applications like blood collection tubes
Multipurpose high-speed centrifuges
devices for a broad range of tube sizes, high variability, big footprint
Ultracentrifuges
analytical and preparative models

Because of the heat generated by air friction (even in ultracentrifuges, where the rotor operates in a good vacuum), and the frequent necessity of maintaining samples at a given temperature, many types of laboratory centrifuges are refrigerated and temperature regulated.

Centrifuge tubes

Centrifuge tubes are precision-made, high-strength tubes of glass or plastic made to fit exactly in rotor cavities. They may vary in capacity from 50 mL down to much smaller capacities used in microcentrifuges used extensively in molecular biology laboratories. Microcentrifuges typically accommodate disposable plastic microcentrifuge tubes with capacities from 250 μL to 2.0 mL.

Glass centrifuge tubes can be used with most solvents, but tend to be more expensive. They can be cleaned like other laboratory glassware, and can be sterilized by autoclaving. Small scratches from careless handling can cause failure under the strong forces imposed during a run. Glass tubes are inserted into soft rubber sleeves to cushion them during runs. Plastic centrifuge tubes, especially tend to be less expensive and, with care, can be just as durable as glass. Water is preferred when plastic centrifuge tubes are used. They are more difficult to clean thoroughly, and are usually inexpensive enough to be considered disposable.

Disposable plastic "microlitre tubes" of 0.5ml to 2ml are commonly used in microcentrifuges. They are molded from a flexible transparent plastic similar to polythene, are semi-conical in shape, with integral, hinged sealing caps.

Larger samples are spun using centrifuge bottles, which range in capacity from 250 to 1000 millilitres. Although some are made of heavy glass, centrifuge bottles are usually made of shatterproof plastics such as polypropylene or polycarbonate. Sealing closures may be used for added leak-proof assurance.

Safety

An Eppendorf laboratory centrifuge Eppendorf centrifuge.JPG
An Eppendorf laboratory centrifuge

The load in a laboratory centrifuge must be carefully balanced. This is achieved by using a combination of samples and balance tubes which all have the same weight or by using various balancing patterns without balance tubes. [2] It is an interesting mathematical problem to solve the balance pattern given n slots and k tubes with the same weight. It is known that the solution exists if and only if both k and n-k can be expressed as a sum of prime factors of n. [3] Small differences in mass of the load can result in a large force imbalance when the rotor is at high speed. This force imbalance strains the spindle and may result in damage to the centrifuge or personal injury. Some centrifuges have an automatic rotor imbalance detection feature that immediately discontinues the run when an imbalance is detected.

Before starting a centrifuge, an accurate check of the rotor and lid locking mechanisms is mandatory. A spinning rotor can cause serious injury if touched. Modern centrifuges generally have features that prevent accidental contact with a moving rotor as the main lid is locked during the run.

Centrifuge rotors have tremendous kinetic energy during high speed rotation. Rotor failure, caused by mechanical stress from the high forces imparted by the motor, can occur due to manufacturing defects, routine wear and tear, or improper use and maintenance. Such a failure can be catastrophic failure, especially with larger centrifuges, and generally results in total destruction of the centrifuge. While centrifuges generally have safety shielding to contain these failures, such shielding may be inadequate, especially in older models, or the entire centrifuge unit may be propelled from its position, resulting in damage to nearby personnel and equipment. Uncontained rotor failures have shattered laboratory windows and destroyed refrigerators and cabinetry. To reduce the risk of rotor failures, centrifuge manufacturers specify operating and maintenance procedures to ensure that rotors are regularly inspected and removed from service or derated (only operated at lower speeds) when they are past their expected lifetime. [4]

Another potential hazard is the aerosolization of hazardous samples during centrifugation. To prevent contamination of the laboratory, rotor lids with special aerosol-tight gaskets are available. The rotor can be loaded with the samples within a hood and the rotor lid fixed on the rotor. Afterwards, the aerosol-tight system of rotor and lid is transferred to the centrifuge. The rotor can then be fixed within the centrifuge without opening the lid. After the run, the entire rotor assembly, including the lid, is removed from the centrifuge to the hood for further steps, maintaining the samples within a closed system.

A ThermoFisher laboratory bench-top centrifuge. Thermo centrifuge.JPG
A ThermoFisher laboratory bench-top centrifuge.

See also

Related Research Articles

<span class="mw-page-title-main">Valve</span> Flow control device

A valve is a device or natural object that regulates, directs or controls the flow of a fluid by opening, closing, or partially obstructing various passageways. Valves are technically fittings, but are usually discussed as a separate category. In an open valve, fluid flows in a direction from higher pressure to lower pressure. The word is derived from the Latin valva, the moving part of a door, in turn from volvere, to turn, roll.

<span class="mw-page-title-main">Test tube</span> Glass or plastic laboratory glassware

A test tube, also known as a culture tube or sample tube, is a common piece of laboratory glassware consisting of a finger-like length of glass or clear plastic tubing, open at the top and closed at the bottom.

<span class="mw-page-title-main">Centrifuge</span> Device using centrifugal force to separate fluids

A centrifuge is a device that uses centrifugal force to subject a specimen to a specified constant force, for example to separate various components of a fluid. This is achieved by spinning the fluid at high speed within a container, thereby separating fluids of different densities or liquids from solids. It works by causing denser substances and particles to move outward in the radial direction. At the same time, objects that are less dense are displaced and moved to the centre. In a laboratory centrifuge that uses sample tubes, the radial acceleration causes denser particles to settle to the bottom of the tube, while low-density substances rise to the top. A centrifuge can be a very effective filter that separates contaminants from the main body of fluid.

<span class="mw-page-title-main">Ultracentrifuge</span> Centrifuge for spinning a rotor at very high speeds

An ultracentrifuge is a centrifuge optimized for spinning a rotor at very high speeds, capable of generating acceleration as high as 1 000 000 g. There are two kinds of ultracentrifuges, the preparative and the analytical ultracentrifuge. Both classes of instruments find important uses in molecular biology, biochemistry, and polymer science.

<span class="mw-page-title-main">Cuvette</span> Small container used in laboratories

In laboratories, a cuvette is a small tube-like container with straight sides and a circular or square cross-section. It is sealed at one end, and made of a clear, transparent material such as plastic, glass, or fused quartz. Cuvettes are designed to hold samples for spectroscopic measurement, where a beam of light is passed through the sample within the cuvette to measure the absorbance, transmittance, fluorescence intensity, fluorescence polarization, or fluorescence lifetime of the sample. This measurement is done with a spectrophotometer.

<span class="mw-page-title-main">Centrifugation</span> Mechanical process

Centrifugation is a mechanical process which involves the use of the centrifugal force to separate particles from a solution according to their size, shape, density, medium viscosity and rotor speed. The denser components of the mixture migrate away from the axis of the centrifuge, while the less dense components of the mixture migrate towards the axis. Chemists and biologists may increase the effective gravitational force of the test tube so that the precipitate (pellet) will travel quickly and fully to the bottom of the tube. The remaining liquid that lies above the precipitate is called a supernatant or supernate.

<span class="mw-page-title-main">Reed switch</span> Electrical switch operated by an applied magnetic field

The reed switch is an electromechanical switch operated by an applied magnetic field. It was invented in 1922 by professor Valentin Kovalenkov at the Petrograd Electrotechnical University, and later evolved at Bell Telephone Laboratories in 1936 by Walter B. Ellwood into the reed relay. In its simplest and most common form, it consists of a pair of ferromagnetic flexible metal contacts in a hermetically sealed glass envelope. The contacts are usually normally open, closing when a magnetic field is present, or they may be normally closed and open when a magnetic field is applied. The switch may be actuated by an electromagnetic coil, making a reed relay, or by bringing a permanent magnet near it. When the magnetic field is removed, the contacts in the reed switch return to their original position. The "reed" is the metal part inside the reed switch envelope that is relatively thin and wide to make it flexible, resembling the reed of a musical instrument. The term "reed" may also include the external wire lead as well as the internal part.

<span class="mw-page-title-main">Differential centrifugation</span> Method of separating particles in a mixture

In biochemistry and cell biology, differential centrifugation is a common procedure used to separate organelles and other sub-cellular particles based on their sedimentation rate. Although often applied in biological analysis, differential centrifugation is a general technique also suitable for crude purification of non-living suspended particles. In a typical case where differential centrifugation is used to analyze cell-biological phenomena, a tissue sample is first lysed to break the cell membranes and release the organelles and cytosol. The lysate is then subjected to repeated centrifugations, where particles that sediment sufficiently quickly at a given centrifugal force for a given time form a compact "pellet" at the bottom of the centrifugation tube.

<span class="mw-page-title-main">Gas centrifuge</span> Device that performs isotope separation of gases

A gas centrifuge is a device that performs isotope separation of gases. A centrifuge relies on the principles of centrifugal force accelerating molecules so that particles of different masses are physically separated in a gradient along the radius of a rotating container. A prominent use of gas centrifuges is for the separation of uranium-235 (235U) from uranium-238 (238U). The gas centrifuge was developed to replace the gaseous diffusion method of uranium-235 extraction. High degrees of separation of these isotopes relies on using many individual centrifuges arranged in series, that achieve successively higher concentrations. This process yields higher concentrations of uranium-235 while using significantly less energy compared to the gaseous diffusion process.

<span class="mw-page-title-main">Watch glass</span> Piece of laboratory glassware

A watch glass is a circular concave piece of glass used in chemistry as a surface to evaporate a liquid, to hold solids while being weighed, for heating a small amount of substance, and as a cover for a beaker. When used to cover beakers, the purpose is generally to prevent dust or other particles from entering the beaker; the watch glass does not completely seal the beaker, so gas exchanges still occur. When used as an evaporation surface, a watch glass allows closer observation of precipitates or crystallization. It can be placed on a surface of contrasting colors to improve the visibility overall. Watch glasses are also sometimes used to cover a glass of whisky, to concentrate the aromas in the glass, and to prevent spills when the whisky is swirled. Watch glasses are named so because they are similar to the glass used for the front of old-fashioned pocket watches. These large watch glasses are occasionally known as clock glasses.

Glass tubes are mainly cylindrical hollow-wares. Their special shape combined with the huge variety of glass types, allows the use of glass tubing in many applications. For example, laboratory glassware, lighting applications, solar thermal systems and pharmaceutical packaging to name the largest.

<span class="mw-page-title-main">Cell disruption</span>

Cell disruption is a method or process for releasing biological molecules from inside a cell.

<span class="mw-page-title-main">Analytical ultracentrifugation</span>

Analytical ultracentrifugation is an analytical technique which combines an ultracentrifuge with optical monitoring systems.

In chemistry, the sedimentation coefficient of a particle characterizes its sedimentation during centrifugation. It is defined as the ratio of a particle's sedimentation velocity to the applied acceleration causing the sedimentation.

<span class="mw-page-title-main">Eye dropper</span> Device used to transfer small quantities of liquids

An eye dropper, also called Pasteur pipette or simply dropper, is a device used to transfer small quantities of liquids. They are used in the laboratory and also to dispense small amounts of liquid medicines. A very common use was to dispense eye drops into the eye. The commonly recognized form is a glass tube tapered to a narrow point and fitted with a rubber bulb at the top, although many styles of both plastic and glass droppers exist. The combination of the pipette and rubber bulb has also been referred to as a teat pipette. The Pasteur pipette name is from the French scientist Louis Pasteur, who used a variant of them extensively during his research. In the past, there was no equipment to transfer a chemical solution without exposing it to the external environment. The hygiene and purity of chemical compounds is necessary for the expected result of each experiment. The eye dropper, both glass and plastic types, can be sterilized and plugged with a rubber bulb at the open end of the pipette preventing any contamination from the atmosphere. Generally, they are considered cheap enough to be disposable, however, so long as the glass point is not chipped, the eye dropper may be washed and reused indefinitely.

A centrifugal evaporator is a device used in chemical and biochemical laboratories for the efficient and gentle evaporation of solvents from many samples at the same time, and samples contained in microtitre plates. If only one sample required evaporation then a rotary evaporator is most often used. The most advanced modern centrifugal evaporators not only concentrate many samples at the same time, they eliminate solvent bumping and can handle solvents with boiling points of up to 220 °C. This is more than adequate for the modern high throughput laboratory.

<span class="mw-page-title-main">Homogenizer</span> Equipment used for the homogenization of various types of material

A homogenizer is a piece of laboratory or industrial equipment used for the homogenization of various types of material, such as tissue, plant, food, soil, and many others. Many different models have been developed using various physical technologies for disruption. The mortar and pestle, already used for thousands of years, is a standard tool even in modern laboratories. More modern solutions are based on blender type instruments, bead mills, ultrasonic treatment, rotor-stator mechanical, high pressure, and many other physical forces. While there are many application overlaps between methods, each homogenization method has distinct advantages and disadvantages.

This is a list of instruments used in general in laboratories, including:

The Zippe-type centrifuge is a gas centrifuge designed to enrich the rare fissile isotope uranium-235 (235U) from the mixture of isotopes found in naturally occurring uranium compounds. The isotopic separation is based on the slight difference in mass of the isotopes. The Zippe design was originally developed in the Soviet Union by a team led by 60 Austrian and German scientists and engineers captured after World War II, working in detention. In the West the type is known by the name of the man who recreated the technology after his return to the West in 1956, based on his recollection of his work in the Soviet program, Gernot Zippe. To the extent that it might be referred to in Soviet/Russian usage by any one person's name, it was known as a Kamenev centrifuge.

A centrifuge is a device that uses centrifugal force to separate its components.